[1]王晗敏,姚月蓉,周欢明,等.IOL MASTER测量低信噪比的高度近视眼人工晶状体度数的精确性分析[J].眼科新进展,2016,36(11):1036-1038.[doi:10.13389/j.cnki.rao.2016.0276]
 WANG Han-Min,YAO Yue-Rong,ZHOU Huan-Ming,et al.Accuracy of IOL Master in measuring IOL power of high myopia with low SNR[J].Recent Advances in Ophthalmology,2016,36(11):1036-1038.[doi:10.13389/j.cnki.rao.2016.0276]
点击复制

IOL MASTER测量低信噪比的高度近视眼人工晶状体度数的精确性分析
分享到:

《眼科新进展》[ISSN:1003-5141/CN:41-1105/R]

卷:
36卷
期数:
2016年11期
页码:
1036-1038
栏目:
应用研究
出版日期:
2016-11-05

文章信息/Info

Title:
Accuracy of IOL Master in measuring IOL power of high myopia with low SNR
作者:
王晗敏姚月蓉周欢明莫利娟张兴儒项敏泓
200062 上海市,上海中医药大学附属普陀医院眼科(王晗敏,周欢明,莫利娟,张兴儒,项敏泓);201900 上海市,宝山区中西医结合眼科(姚月蓉)
Author(s):
WANG Han-Min YAO Yue-Rong ZHOU Huan-Ming MO Li-Juan ZHANG Xing-RuXIANG Min-Hong
Department of Ophthatmology , Putuo Hospital , Shanghai University of Traditional Chinese Medicine ( WANG Han-Min, ZHOU Huan-Ming, MO Li-Juan , ZHANG Xing-Ru , XIANG Min-Hong ) , Shanghai 200062 . China ; Department of Ophthalmology, Baoshan District Traditional Chinese and Western Medicine Hospital ( YAO Yue-Rong ) , Shanghai 201900 , China
关键词:
IOL MASTERA超信噪比高度近视
Keywords:
IOL Master A-scan signal to noise ratio high myopia
分类号:
R778.1
DOI:
10.13389/j.cnki.rao.2016.0276
文献标志码:
A
摘要:
目的 比较A超与光学相干生物测量仪(IOLMASTER)测量低信噪比(signaltonoiseratio,SNR)高度近视眼人工晶状体(intraocularlens,IOL)度数的精确性。方法 对2015年3月至2016年2月间在我院行白内障超声乳化联合IOL植入手术并且眼轴长度大于26mm的高度近视患者57例(69眼)进行回顾性研究。SNR≤2组11眼、SNR>2~5组23眼、SNR>5~10组20眼,无法用IOLMASTER检出的病例设为A超测量组15眼。分别用IOLMASTER或接触式A超测算度数,计算公式采用HOLLADAY-1公式,并按各自方法植入同一类型IOL,术后3个月分别计算各组的平均绝对屈光误差值。结果 SNR>5~10组、SNR>2~5组、SNR≤2组、A超测量组的绝对屈光误差值分别为(0.44±0.18)D、(0.46±0.26)D、(0.69±0.22)D、(0.81±0.29)D,其中SNR>2~5组及SNR>5~10组分别与其他2组比较,差异均有统计学意义(均为P<0.05),而SNR>2~5组与SNR>5~10组以及SNR≤2组与A超测量组之间差异均无统计学意义(均为P>0.05)。结论对于低SNR的高度近视眼患者,采用IOLMASTER测算IOL度数比传统A超方式具有更高的精确性。
Abstract:
Objective To compare the accuracy of A-scan and IOL Master in measuring intraocular lens ( IOL) power of high myopia with low signal to noise ratio ( SNR) . Methods From March 2015 to February 2016,69 eyes of 57 patients of high myopia with axial length greater than 26 mm who received phacoemulsification with IOL implantation in our hospital were enrolled. According to the SNR , eyes were assigned to SNR¡?2 goup , >2 -5 group and > 5 - 10 group , and those eyes which could not be measured by IOL Master were assigned to A-scan group. A1l eyes were implanted same type of artificial IOL which were calculated through HOLLADAY-I. The mean absolute error was evaluated at 3 months after the operation. Results The mean absolute error in SNR > 5 - 10 group , > 2 - 5 goup ,¡Ü2 group and contact A-scan group were (0. 44 +0. 18 ) D, ( 0. 46 +0. 26) D. (0. 69 + 0. 22 ) D and ( 0. 81 + 0. 29 ) D, respectively. The SNR > 2 - 5 group and SNR > 5 - 10 group were respectively compared with the other two groups,the differences were statistically siguficant ( all P < 0. 05 ) . There was no siguficant difference between SNR > 2 - 5 group and SNR > 5 - 10 ~oup , or between SNR¡?2 group and contact A-scan group ( all P > 0. 05 ) . Conclusion The calculation of IOL power by IOL Master is more accurate than that by contact A-scan for high myopia with low SNR.

相似文献/References:

[1]胡俊 李学喜 林巧雅 叶瑞珍 许敏.不同仪器测量角膜直径和角膜曲率的一致性评价[J].眼科新进展,2012,32(4):000.
[2]李翔 庞辰久 彭海鹰 戴丽娟 孟志红 鲁传琴.Orbscan-II、IOL Master和Atlas角膜地形图测量角膜直径的比较[J].眼科新进展,2012,32(11):000.
[3]陈晓 李舒茵 牛超 李建新 陈慷.先天性视网膜劈裂症的影像学观察[J].眼科新进展,2012,32(11):000.
[4]孙靖 梁四妥 田芳 苏龙 张红.硅油眼眼轴长度的测量研究[J].眼科新进展,2012,32(11):000.
[5]吴敏,肖丽波,胡竹林,等.IOL-Master在OxaneHD填充眼眼轴测量中的应用[J].眼科新进展,2014,34(11):1049.[doi:10.13389/j.cnki.rao.2014.0290]
 WU Min,XIAO Li-Bo,HU Zhu-Lin,et al.Outcome analysis of IOL-Master in measuring axial length of Oxane Hd filled eyes[J].Recent Advances in Ophthalmology,2014,34(11):1049.[doi:10.13389/j.cnki.rao.2014.0290]
[6]罗书科,叶祖科,尹小芳,等.散瞳对Lenstar和A超测量白内障患者眼生物学参数的影响[J].眼科新进展,2016,36(2):151.[doi:10.13389/j.cnki.rao.2016.0041]
 LUO Shu-Ke,YE Zu-Ke,YIN Xiao-Fang,et al.Effects of pupil dilation on ocular biometric parameters measured by Lenstar and A-scan in cataract patients[J].Recent Advances in Ophthalmology,2016,36(11):151.[doi:10.13389/j.cnki.rao.2016.0041]
[7]孙飒,郭海科.AL-Scan、IOL-Master和A超测量白内障患者眼轴长度的准确性和一致性分析[J].眼科新进展,2016,36(9):876.[doi:10.13389/j.cnki.rao.2016.0235]
 SUN Sa,GUO Hai-Ke.Accuracy and consistency research of AL-Scan, IOL-Master and A-type ultrasound scan for axial length measurement in cataract patients[J].Recent Advances in Ophthalmology,2016,36(11):876.[doi:10.13389/j.cnki.rao.2016.0235]